With the weight still on the pan, the
display will now show the weight that is
7
calibrated. Check calibration to ensure the
scale is reading correct
INTERNAL SERVICE CALIBRATION
7.2
Internal calibration also referred as service calibration requires
opening the machine and changing the jumpers. It is advisable to
do this step only if you have required assistance or the service
personnel assisting you to do this.
STEP
Open the 4 screws from the bottom of the
machine and change the calibration
jumper marked CAL. Now place back the
cover, ABS pan and the SS pan and switch
1
on the machine. Make sure there is
nothing kept on the weighing pan. At this
moment do not put back the screws,
simply place the cover on the machine.
To start calibration turn the scale off and
then turn it on. During the count down,
2
press and hold down the Zero button (do
not release until the display shows "CAL"
Make sure there is no weight on the pan.
Press Zero / Enter key and the display
3
shows "noLoAd" momentarily and shows
the counts at zero level.
PROCEDURE
14
NANO Series Precision Balance
DISPLAY