Olympus BHSP Kullanım Kılavuzu - Sayfa 11

Mikroskop Olympus BHSP için çevrimiçi göz atın veya pdf Kullanım Kılavuzu indirin. Olympus BHSP 22 sayfaları. Bhs-p series polarizing microscope

Olympus BHSP Kullanım Kılavuzu
Summary of Putting the Microscope in Operation
Model BHSP
Match the line voltage selector switch t o local mains voltage (see page
10).
Switch on the light source.
Place a specimen slide on the stage.
Remove the Bertrand lens and analyzer from the light path.
Coarse focus with the
10X
objective.
Make interpupillary and diopter adjustments (page
10).
Set the analyzer to optimum extinction position (page
11).
Center the condenser (page
12).
Center the stage (page
13).
J
.
Center objectives other than
10X
(page
13).
K .
Swing in the desired objective.
L.
Set the condenser, analyzer and Bertrand lens correctly according t o your microscopic
2
purpose (pages
15
and
16)
@
+
M.
Fine focus.
0
N.
Adjust aperture iris diaphragm and field iris diaphragm (page
13).
/
Adjustment of Illumination System
I
1
orthosco~ic
/
4X
to
loox
I
observation
OUT
I
OUT
1
Microscopic
application
Generally for biological use, however, remove the analyzer, Bertrand lens and test plates
Objective
Conoscopic
observation
from the light path.
I
*
Cut off this page at dotted line and put it on the wall near the microscope for use as a
reminder of microscopic procedure.
i
Bertrand lens
in intermediate
polarizing at-
tachment
20X
to
100X
OLYMPUS
Condenser
top lens
IN
I N