Olympus BX-RLA2 Talimatlar Kılavuzu - Sayfa 17

Mikroskop Olympus BX-RLA2 için çevrimiçi göz atın veya pdf Talimatlar Kılavuzu indirin. Olympus BX-RLA2 32 sayfaları. Reflected light brightfield/darkfield illuminator

Olympus BX-RLA2 Talimatlar Kılavuzu
OBSERVATION METHODS (Using BX-RLA2)
6-1 Reflected Light Brightfield Observation
See "REFLECTED LIGHT BRIGHTFIELD/DARKFIELD OBSERVATION PROCEDURE" on page 5.
6-2 Reflected Light Nomarski DIC (Differential Interference Contrast) Observation
# The performance of polarizer may deteriorate when it has been exposed to light for a long period (about continu-
ous 2000 hours). If this happens, replace the polarizer.
# When performing sensitive color observation using the U-DICRH DIC slider, combine the U-POTP3 polarizer.
#When using the high-intensity light source, be sure to use the U-25L42 filter to prevent the polarizer from being burnt.
@
Fig. 10
1
Setting the Analyzer and Polarizer
# In the current step, do not engage the DIC slider in the light path.
1. Bring the specimen in approximate focusing using the 10X or 20X objec-
²
tive.
2. Remove the cover to enable installation of the U-AN360-3 analyzer. Then
put the analyzer @ in the insertion slot.
3. Insert the U-PO3 or U-POTP3 polarizer ² so that the surface with indica-
tion comes on the front, and engage the polarizer in the light path.
³
4. Rotate the analyzer rotating dial ³ to find the position where the field of
view is darkest.
}An approximate cross-Nicol position can be obtained by setting the index
on the dial ³ on the outer side. Fine-adjust the dial by rotating it near this
position to find the position where the field of view is darkest.
BX-RLA2
(Fig. 10)
14