Olympus BX41-P Talimatlar Kılavuzu - Sayfa 19

Mikroskop Olympus BX41-P için çevrimiçi göz atın veya pdf Talimatlar Kılavuzu indirin. Olympus BX41-P 36 sayfaları. Polarizing microscope
Ayrıca Olympus BX41-P için: Talimatlar Kılavuzu (36 sayfalar)

Olympus BX41-P Talimatlar Kılavuzu

POLARIZED LIGHT OBSERVATION

5-1 Adjustments Before Observation

In polarized light microscopy ultimate performance is not obtainable unless optical adjustments are made correctly.
Always performed the following adjustments before observation. Remove the quarter wave plate and sensitive tint plate
from the light path.
Fig. 19
1

Adjusting the Optical Axis

When Using the Intermediate Attachment for orthoscopic observation
(U-OPA) (Figs. 19 & 20)
³
²
CAUTION
Make sure the 10X objective is inserted into the primary hole on
the centerable revolving nosepiece.
@
1. Fully insert the alignment polarizer @, provided with the U-OPA into
the test plate adapter (U-TAD). Tighten its clamping knob. (Fig. 19)
2. Remove the condenser.
3. Insert the rotatable analyzer (U-AN360P-2) ² into the analyzer slot on
the intermediate attachment. Engage the analyzer, loosen the clamping
screw and set the analyzer's rotation dial ³ to 0°. (Fig. 19)
BX53/51/41-P
16