HALOGEN LAMP SOCKET INSPECTION SHEET

{ For safe use of the lamp socket, we recommend performing the following inspection periodically (every time you replace
the lamp bulb and at least every 6 months).
{ The table below identified the check items to be observed. Put (X) if not applicable or ( ) if applicable.
{ If there is any ( ) mark noted, immediately stop use of the product, and consult Olympus for detailed inspections or replace
the lamp soket.
{ If you detect an abnormality other than that listed below or with other Olympus product, also stop the use of the product and
contact Olympus for detailed inspections.
{Note that the service, replacement and detailed inspections are charged after expiration of the warranty period.
1. More than 8 years have passed since original purchase or the total power
ON time exceeds 20,000 hours.
2. Lamp does not light sometimes even though the main switch is set to on.
3. Illumination flickers when you move the lamp cable or lamp socket.
4. Scorching or burning odor is produced during use.
5. Illumination still flickers after replacement with a new lamp bulb.
6. Deformation, backlash, or looseness, etc. when you assemble the lamp socket.
7. Extreme discoloration of the lamp socket connection terminal or lamp socket
lamp bulb mount.
8. Discoloration, deformation or cracking of the lamp socket.
9. Melting, crack, deformation or solidification of the lamp cable or a wiring part.
10. Increased frequency of servicing compared to similar devices put into use at
the same time as the lamp socket.
* When the Check Result columns become insufficient, copy this sheet.
Check items
If you have any questions, please contact Olympus.
Check results (Date)
/
/