6.
Loosen the clamping screws inside four holes (G) on top of the Autofocus Sensing Unit BXC-FSU, then move BXC-FSU
in all directions to place both the center of the objective pupil and the center of the CT cross within one scale. When
this is achieved, securely tighten the clamping screws.
NOTE:
Do not loosen the clamping screws too much.
7.
Remove the standard cylindrical shaft and the CT, then cover four holes (E) on top of the Autofocus Sensing Unit BXC-
FSU with the provided seals.

3.4 Installing the control box

When securing the control box to the floor, wall or ceiling, attach it with your screws.
Positions of screw holes:
Recommended screw: M4, L: 8 mm or longer
Washer: M4
1.
Remove four rubber feet from the control box using a tool such as tweezers.
2.
Insert your washers between the screw holes and the screws and tighten four screws.
NOTE:
Do not tighten the screws too much. Doing so may damage the product.
9
3. Setup procedures
3.4 Installing the control box