Olympus CH2 Series Onarım Kılavuzu - Sayfa 25
Mikroskop Olympus CH2 Series için çevrimiçi göz atın veya pdf Onarım Kılavuzu indirin. Olympus CH2 Series 42 sayfaları. Biological mlcroscope
l-2-6.
l-2-7.
Put the arm in its original
upright
position
and adjust the pinion shaft and the nut.
1. Gradually
tighten
the nut while confirm-
ing
that
there
is no excess friction,
uneven movement
or play in the thrust
direction.
2. If the nut is tightened
too much it may
cause excess friction.
Be careful.
Apply
glue (OTI 027) on the threads
nut.
of the
2.
Assembling
the upper stopper of coarse focus-
ing unit
2-l.
Mount
the clu
bly.
*Confirm
that
undamaged.
tch plate on the pinion
assem-
the 1 plate is free of grease and
-26-