Camera Unit
|
@
²
³
Fig. 37
2
Mounting the Camera Unit
1. Loosen the clamping screw ² of the PM-ADF NFK photo eyepiece
adapter @.
2. Attach the adapter @ to the CH3-TR45 trinocular tube ³ and tighten the
screw ².
3. Insert 3.3X or 5X NFK photo eyepiece | into the adapter @.
4. Mount the camera unit directly on the adapter @. Make sure the viewfinder
on the camera unit faces sideways. If it faces to the front, the eyepieces
will get in your way when you try to look into the camera unit's viewfinder.
If the viewfinder of the PM30 faces sideways, it will interfere with the
connection cable. To avoid this, mount the PM30 on the adapter at about
a 30° angle with respect to the eyepieces.
When using the OM camera back
Mount the MOM-L photomicro adapter onto the OM camera back in the
same way as you would an interchangeable lens.
· Use the 3.3X or 5X NFK photo eyepiece.
· Look into the viewfinder on the camera back to focus on the specimen.
· Photomicrographic magnification
= Objective magnification X Photo eyepiece magnification.
# When taking microscopic pictures with an SLR camera back, mirror-
lockup shock occurs. To minimize the camera-shake associated with
this shock, select a longer exposure time to obtain clearer pictures.
If it is not possible to select a longer exposure time, use the ND filter
to reduce the light intensity.
3
Focusing
Focusing is not coupled between the observation tube´s eyepiece and
the film surface. Be sure to use the camera unit´s viewfinder to focus on
the specimen when taking a photograph. For details, refer to the camera
unit's instruction manual.
4
Color Temperature Adjustment
}When taking color pictures using daylight type film, follow the proce-
dures below.
1. Mount the 45LBD-2N color temperature conversion filter in the filter holder.
2. Turn the brightness adjustment control to set it at its maximum position.
Illumination close to daylight will be obtained.
CK30/CK40
(Fig. 37)
24