Olympus CKX31 Onarım Kılavuzu - Sayfa 40
Mikroskop Olympus CKX31 için çevrimiçi göz atın veya pdf Onarım Kılavuzu indirin. Olympus CKX31 43 sayfaları. Culture microscopes
Ayrıca Olympus CKX31 için: Genel Bakış (7 sayfalar), Broşür ve Teknik Özellikler (4 sayfalar)
CKX31/CKX41
Parfocality adjustment amount (CKX41: reference value)
Reading value
(before adjustment)
AB129900 AB130000 AB130200 AE092100 AE092200 AE092300
2.0
1.9
1.8
1.7
1.6
1.5
1.4
1.3
1.2
1.1
1.0
0.9
0.8
0.7
0.6
0.5
0.4
0.3
0.2
0.1
0.0
-0.1
-0.2
-0.3
-0.4
-0.5
-0.6
-0.7
-0.8
-0.9
-1.0
-1.1
-1.2
-1.3
-1.4
-1.5
-1.6
-1.7
-1.8
-1.9
-2.0
Reading value (before adjustment): Reading value of parfocality difference on the helicoid scale of standard eyepiece
Reading value (after adjustment): Theoretical reading value of parfocality difference after adjustment
Ex.) Parfocality differnce is 2.0 mm: when the total thickness (1.05mm) of washers, (0.05mm: 1pc.) and ( 1mm: 1 pc.),
* Numeric character (1 or 2) in the column of washer selection shows the number of washers.
D. REPAIR PROCEDURE
Washer selection
0.05
0.1
0.3
1
1
1
2
1
2
2
2
2
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
are inserted , parfocality difference becomes -0.03mm reading value.
< standard : 0.7 >
0.7
0.5
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
D-13
Unit: mm
Total
Reading value
washer
(after adjustment)
thickness
1.0
1.05
-0.03
1
1.05
1
-0.13
1
1.00
0.06
1
1.00
-0.04
-0.14
1
1.00
0.95
0.05
-0.05
0.95
0.90
0.14
0.90
0.04
0.90
-0.06
0.85
0.13
0.03
0.85
-0.07
0.85
0.12
0.80
0.80
0.02
0.80
-0.08
0.75
0.11
0.75
0.01
0.75
-0.09
0.70
0.10
0.70
0.00
0.70
-0.10
0.65
0.09
0.65
-0.01
0.65
-0.11
0.60
0.08
0.60
-0.02
0.60
-0.12
0.55
0.07
0.55
-0.03
0.55
-0.13
0.50
0.06
0.50
-0.04
0.50
-0.14
0.45
0.05
0.45
-0.05
0.40
0.14
0.40
0.04
0.40
-0.06
0.35
0.13
0.35
0.03