Olympus CKX41 Onarım Kılavuzu - Sayfa 19
Mikroskop Olympus CKX41 için çevrimiçi göz atın veya pdf Onarım Kılavuzu indirin. Olympus CKX41 43 sayfaları. Reflected fluorescence system
Ayrıca Olympus CKX41 için: Genel Bakış (7 sayfalar), Broşür ve Teknik Özellikler (4 sayfalar), Talimatlar Kılavuzu (36 sayfalar)
CKX31/CKX41
No.
Parts name
(A)
COVER
(B)
COVER
(C)
COVER
(D)
BINOCULAR TUBE CUKK3X6SA, 4pcs. (*4)
(E)
INDICATOR PLATE
(F)
DIOPTER RING
(G)
HELICOID ASS'Y
(H)
SLEEVE ASS'Y
(I)
PRISM ASS'Y
(J)
PRISM ASS'Y
(K)
TORQUE PLATE
(L)
SPRING WASHER
(M)
WASHER
(N)
MOUNT
(O)
RING
(P)
PRISM MOUNT-R
(Q)
WASHER
(R)
PRISM MOUNT-L
(S)
RING
(T)
SHAFT
(U)
FIXING PLATE
(V)
PRISM
* In this manual, left ( right) sleeve indicates the left (right) side with the observed condition.
C. DISASSEMBLY AND ASSEMBLY PROCEDURES
Screw
3PUTB2X4SB, 3pcs. (*1)
3PUTB2X4SB, 3pcs. (*2)
CUTB3X10SA, 2pcs. (*3)
ACU3X3SA, 2pc. (*5)
HU3X4SA, 1pc. (*6)
CUK3X6SB, 2pcs. (*7)
CUK3X6SB, 2pcs. (*8)
CUK3X8SA, 2pcs. (*9)
CUK3X8SA, 2pcs. (*10)
AWU3X4SA, 2pcs. (*11)
CUK3X6SA, 2pcs. (*12)
C-8
Grease Adhesive
Attach it with double coated
adhesive tape.
Since the mounting procedure is
the samet as that of CKX31,
refer to C-6.
Refer to D-2.
Left/right optical axis:
Follow the adjustment procedure
on D-8.
Revolving axis:
Follow the adjustment procedure
on D-8.
Push it in the arrow directions
during assembling.
Exit pupil center:
Follow the adjustment procedure
on D-8.
Same as the above
Tension: standard 5-15N
Since the adjustment is the same
as that of CKX31, refer to C-6.
OT3189
OT3189
OT3189
OT3189
Apply grease to the contacting
surface only.
Absolute optical axis:
Follow the adjustment procedure
on D-8.
Remark
(OT3223)