2 Replacement of Cubes

2-1 Replacing the DM Cube

The DM cube is used to connect the light path of the optional FV10-ASU Auxiliary Scan
Unit or FV10-OPD Non-confocal Point Detector with that of the scan unit, and should be
selected according to the observation method.

2-1-1 With the FV10-ASU

1.
2.
3.

Replacement of Cubes /

Set the light path of the scan unit to the LSM light path. (This can be done with the
FLUOVIEW software. For details, refer to the User's Manual for the FLUOVIEW
software.)
Loosen the four cover clamping knobs
the scan unit, and remove the cover
Using an Allen screwdriver, loosen the screw
move the guide lock plate in the direction of the arrow, engage it with the pin
below the guide, and tighten the screw
Before moving
Replacing the DM Cube
on the lower part of the right side panel of
.
retaining the guide lock plate
again to lock the guide.
After moving