Olympus U-BI30-2 Talimatlar Kılavuzu - Sayfa 31

Mikroskop Olympus U-BI30-2 için çevrimiçi göz atın veya pdf Talimatlar Kılavuzu indirin. Olympus U-BI30-2 36 sayfaları. Motorized illumination system with motorized focusing unit

Olympus U-BI30-2 Talimatlar Kılavuzu

8-3 Detailed Assembly Procedure

@
²
Fig. 22
³
Fig. 23
5
6
4
7
Fig. 24
3
Attaching the Halogen Bulb
}Use only the designated bulb 12V100WHAL-L (PHILIPS 7724) or
12V50WHAL-L (LIFE JC).
1. Fully loosen the lamp housing clamping screw 1 on top of the lamp
housing using the Allen screwdriver provided with the illuminator.
2. Lift the lamp housing cover 2 upward to remove it.
3. Turn the lamp socket 3 by 90° in the direction indicated by the arrow.
4. Holding the bulb 5 with gloves or a piece of gauze, depress the bulb
clamping lever 4 and insert the bulb pins 6 fully into the pin sections 7
on the lamp socket.
Gently release the bulb clamping lever to the original position to secure
the bulb.
!To prevent reduced bulb life or cracking, do not touch the bulb with
bare hands. If fingerprints are accidentally left on the bulb, wipe the
bulb with a soft cloth.
5. Slide the lamp housing cover onto the housing base from the above.
Tighten the clamping screw 1 while pressing downward on the cover.
(Fig. 22)
!Caution for Bulb Replacement During Use or Right After Use
The bulb, lamp housing and areas near these will be extremely hot
during and right after use.
Set the main switch to "
the wall outlet, then allow the old bulb and lamp housing to cool
before replacing the bulb with a new of the designated type.
(Figs. 22 to 24)
" (OFF), disconnect the power cord from
BXFMA
27