Olympus U-IFFH Talimatlar Kılavuzu - Sayfa 5
Mikroskop Olympus U-IFFH için çevrimiçi göz atın veya pdf Talimatlar Kılavuzu indirin. Olympus U-IFFH 36 sayfaları. Motorized illumination system with motorized focusing unit
IMPORTANT
The BXFMA-F motorized illuminator with motorized focusing unit can be installed in a user's system in
combination with the BX-UCB control box as well as a PC (running the BX2-BSW software) or the
U-HSTR2, U-FH focus adjustment knob unit and U-IFFH focus adjustment knob interface.
The instructions for the BX/UCB/U-HSTR2 and software (BX2-BSW, commands) are provided in separate
instruction manuals. Please also refer to these manuals as required.
Motorized Controls
Configuration of Instruction Manuals
This motorized illuminator can manifest its functions when it is combined with other modules. The instructions
manuals for the following modules are prepared separately to the present instruction manual. Please read the
manuals according to your own system.
Manual Name
BX-UCB/U-HSTR2
BX2 Software for PC (CD-ROM)
BX2-BSW Ver. 03.01 or higher
TH4
U-FWT/FWR/FWO
Precautions When Unpacking the Microscope
}To prevent deterioration in performance during transport, the illuminator is shipped with transport lock applied to the
focusing unit, observation method switch and aperture iris diaphragm (AS) mechanism, Be sure to release the transport
lock of these three modules before assembly, and engage the transport lock before the next transport of the illuminator.
Fig. 1
²
³
Fig. 2
· BF (brightfield)/DF (darkfield) switching operation
· AS (aperture iris diaphragm) opening/closing operation
· Focus adjustment operation
Functions of the Control Box (incorporating the power supply)
and Hand Switch
Methods of PC control of functions
# Normal operation is not available unless the specified software is used.
External halogen lamp power supply
Motorized filter wheels (Only the U-FWR can be used in this system. )
Releasing the Transport Lock of the Observation Method (BF/DF)
Switch Mechanism (Fig. 1)
Turn the lock knob 1 counterclockwise to loosen it, and remove it together
with the caution tag.
}Retain the lock knob carefully because it will be used again the next time
the equipment is transported.
To set transport lock:
1. Switch the observation method to DF (reflected light darkfield).
2. Turn the lock knob all the way clockwise.
@
Releasing the Transport Lock of the Focusing Unit Mechanism (Fig. 2)
1. Turn the two lock knobs 2 counterclockwise to loosen and remove them.
2. Similarly to step 1, turn the two lock screws 3 to remove them.
}Retain the lock knobs and lock screws carefully because they will be
²
used again the next time the equipment is transported.
To set transport lock:
1. Raise the focusing unit to the higher limit position.
³
2. Turn the two lock screws all the way clockwise.
3. Similarly to step 2, turn and tighten the too lock knobs.
Main Contents
BXFMA
1