Olympus U-LGPS Talimatlar Kılavuzu - Sayfa 16
Mikroskop Olympus U-LGPS için çevrimiçi göz atın veya pdf Talimatlar Kılavuzu indirin. Olympus U-LGPS 24 sayfaları. Led and ldp light source
D
B
D
12
3
Adjusting the brightness
Rotate the light intensity adjustment dial E to adjust the light
1
intensity.
· Rotating the light intensity adjustment dial E allows you to
TIP
E
· The larger the numerical characters on the counter D
When using U-LGPS control software:
· You cannot adjust the light intensity using the light intensity
4
Pilot lamp color and counter display
The relation between the color of the pilot lamp B and the display
on the counter D are listed as follows.
Color
Green Lamp is OFF
Blue
Red
* The procedure to recover from the error is described below.
Display
c06
Other than
c06
If you cannot recover from the error according to the above
procedure, contact Olympus for assistance.
When using U-LGPS control software:
The color of the pilot lamp B changes as the table shown above,
and the display of the counter D always shows 000.
change the numerical characters (Min: 010, Max: 100) on the
counter D in increments of 5.
become, the brighter the lamp becomes.
adjustment dial E . Adjust the light intensity on the U-LGPS
control software.
Pilot lamp B
Status
Lamp is ON
Error in U-LGPS (Lamp
turns OFF automatically)
Recovery procedure
Insert the liquid light guide into the liquid
light guide insertion port of U-LGPS properly.
Set the power switch A to
for a while and turn ON the lamp.
Display on counter D
Display
Description
010-100
Brightness of the
lamp when the
lamp is ON
c01-c23 *
Details of the
error in U-LGPS
Page
15
11
(OFF), wait