STEP 4
Anchor the jaw insert with your left hand then rotate the shaft counter-clockwise
a half-turn to dislodge the jaw insert from the insulated shaft.
STEP 5
Remove the jaw insert from the insulated shaft.
NOTE: Push down on the red release button if disassembling bipolar instruments.
STEP 6
Remove the red sealing cap on the insulated shaft and the black sealing cap from
the handle. Prepare for cleaning and sterilization.