Olympus MU-1 Bakım Kılavuzu - Sayfa 7
Tıbbi Ekipman Olympus MU-1 için çevrimiçi göz atın veya pdf Bakım Kılavuzu indirin. Olympus MU-1 28 sayfaları. Maintenance unit
Ayrıca Olympus MU-1 için: Talimatlar Kılavuzu (36 sayfalar)
Chapter 1: Product Outline
This product is used in combination with the AW channel cleaning adapter, leak tester and
chemical tank to provide the following functions.
(1)
This product can be combined with the AW channel cleaning adapter to drain water from
the OES scope and EVIS scope air and water channel.
(2)
This product can be combined with the leak tester to test OES scopes and EVIS scopes
for water leaks by applying air pressure to the interior of the scopes and then submerging
them in water to look for an air bubbles.
(3)
This product can be combined with the AW channel cleaning adapter and chemical tank to
feed rubbing alcohol into the OES scope and EVIS scope air and water channel in order to
dry them.
(1)
This product simplifies draining water from the OES scope and EVIS scope air and water
pipelines. (AW channel cleaning adapter)
(2)
This product simplifies the water leak tests for the OES scope and EVIS scope. (Using
the leak tester.)
(3)
The product can be installed on a wall near a sink with the included hanger.
(4)
This product can be installed on the left or right side of a sink.
(5)
This product simplifies feeding rubbing alcohol into the OES scope and EVIS scope air
and water channel in order to dry them. (AW tube cleaning adapter, chemical tank)
ISSUE6
1. Product Overview
2. Features
1-1
MU-1
Product Outline