4.4 Focusing

The R1 provides two focusing options: autofocusing and manual focusing. The R1 has an autofocusing
algorithm optimized for cell counting that works in tandem with a novel focusing mechanism that rapidly obtains
the Z position of the sample by the application of a small voltage to a liquid lens. The elimination of mechanical
parts removes noise and significantly reduces the need for servicing.

4.4.1 Autofocusing

Press the circle next to [Autofocused Counting]. The circle will turn blue when the autofocus is activated.

4.4.2 Manual Focusing

Users can adjust the focus manually by simply pressing the [Focus] arrow heads (up or down) with the
autofocus function on or off.

4.5 Cell Counting

Use a finger or a stylus to navigate the image. The red outer box in the top left corner of the image represents
the entire counting area and the inner box is the current field of view. The size and location of the inner box will
change with the magnification and movement of the screen
Press the magnifier button to zoom in and out of the image.