AccuMed CMS-50D Manuel - Sayfa 7

Tıbbi Ekipman AccuMed CMS-50D için çevrimiçi göz atın veya pdf Manuel indirin. AccuMed CMS-50D 14 sayfaları. Fingertip pulse oximeter
Ayrıca AccuMed CMS-50D için: Kullanıcı Kılavuzu (2 sayfalar)

D. The SpO
sensor should not be used at a location or limb tied with arterial canal or blood
2
pressure cuff or receiving intravenous injection.
E.
Make sure the optical path is free from any optical obstacles like rubberized fabric, otherwise it
may result in venous pulsation and inaccurate measure of SpO
F.
Excessive ambient light may affect the measuring result. It includes fluorescent lamp, dual ruby
light, infrared heater, direct sunlight and etc.
G. Strenuous action of the testee or extreme electrosurgical interference may also affect the
accuracy.
H. Testee can not use enamel or other makeup.
I.
Please clean and disinfect the device after operating according to the user manual.

3.3. Clinical restrictions

A. As the measure is taken on the basis of arteriole pulse, substantial pulsating blood flow of the
testee is required. For a testee with weak pulse due to shock, low ambient/body temperature,
major bleeding, or use of vascular contracting drug, the SpO
In this case, the measurement will be more sensitive to interference.
B. For those with a substantial amount of staining dilution drug (such as methylene blue, indigo
green and acid indigo blue), or carbon monoxide hemoglobin (COHb), or methionine (Me+Hb)
or thiosalicylic hemoglobin, and some with icterus problem, the SpO
monitor may be inaccurate.
C. The drugs like dopamine, procaine, prilocaine, lidocaine and butacaine may also be a major
factor blamed for serious error of SpO
D. The SpO
value serves only as a reference value for judgement of anemic anoxia and toxic
2
anoxia, some patients with serious anemia may also report good SpO

4. Technical specifications

A. Display mode: double Color OLED display
SpO
measuring range:
2
Pulse rate measuring range: 30bpm~240bpm
Pulse waveform display:
Battery indication:
B. Power supply requirement:
1.5V (AAA size) alkaline batteries × 2
Adaptable range: 2.6V~3.6V
measure.
2
35%~100%
bar graph display
OLED Signal
4
.
2
waveform (PLETH) will decrease.
2
determination by this
2
measurement.
2