Any solution with a high or low pH may damage the instrument. Enzyme concentration
should follow manufacturer recommendations. Do not use dish soap, iodine, bleach, presept,
cold-soak solution, chlorhexidine-based solution, laundry soap or surgeons hand scrub.
2. Scrub chuck adapter, linen, buckles and chuck key in the enzymatic solution using a soft-
bristled, non-abrasive brush under the surface of the water to minimize splashing. Scrub
the inside of the chuck and shaft using a channel cleaning brush. Pay special attention to
crevasses on the buckles and outside of the metal chuck adapter. Remove all visible soil.
Metal brushes or scouring pads will damage the linen.
3. Open and close the chuck while flushing
irrigation through the chuck in order to
remove any residual soil. When done,
leave the chuck in fully open position.
4. Carefully inspect for any signs of residual
soils. Repeat steps 2 and 3 until no
residuals remain.
5. Rinse in water and allow device to drain.
Remove excess moisture with lint-free,
soft absorbent towel. Allow device to dry
completely before sterilization. Remove
excess moisture with lint-free, soft
absorbent towel to speed up drying. Do
not wring or twist device. Hang cover in
drying cabinet, or on a drying rack in front
of a fan.
Autoclaving the cover while wet will reduce
its expected lifetime.