Accuris ASPIRE Kullanıcı Kılavuzu - Sayfa 3

Laboratuvar Ekipmanları Accuris ASPIRE için çevrimiçi göz atın veya pdf Kullanıcı Kılavuzu indirin. Accuris ASPIRE 8 sayfaları. Laboratory aspirator

6. For repairs, contact Benchmark Scientific technical services at 908-769-5555 and always use original spare
parts and accessories.
above can jeopardise the safety of the user of the device;
7. This laboratory instrument must be used exclusively for its intended applications as described in this
manual.
8. Any different use could be dangerous; the manufacturer will not be responsible for damage due to improper
use or connection to an electrical system not complying with current regulations;
9. Internal electrical and mechanical parts are not designed to be repaired by customers or end-users. Do not
open the pump unit, do not mishandle the electric / mechanical parts. Always contact Benchmark Scientific for
technical assistance
10. Using the device in environmental conditions different than those indicated in this manual may be dangerous to
the user and may cause damage to the instrument.
MODEL
VOLTAGE
POWER CONSUMPTION
FUSE
MAXIMUM SUCTION PRESSURE (without jar)
MAXIMUM SUCTION FLOW (without bottle / filter)
WEIGHT
DIMENSIONS
SICILICONE TUBE SIZE
ACCURANCY OF VACUUM INDICATOR
AMBIENT OPERATING CONDITIONS
STORAGE AND TRASPORT CONDITIONS

TECHNICAL SPECIFICATIONS

ACCURIS ASPIRE LABORATORY ASPIRATOR
Adjustable from -75kPa (-0.75bar) to -10kPa (-0.10 bar)
Room temperature:
Room humidity percentage:
Altitude:
Room temperature:
Room humidity percentage:
SYMBOLS
Storage/transport temperature: -40 ÷ 70°C
3
115V,50/60Hz or 230V, 50/60Hz
105VA
/
184VA
F 1 x 4A 250V / F 1 x 1.6A 250V
-75kPa (-0.75 Bar)
15 l/min
4.8 Lbs / 2.2 Kg
250mm x 190mm x 160 mm
Ø 6 x 10 mm
± 5%
5 to 35°C
30 to 75% RH
0 to 2000m
-40 to 70°C
10 to 100% RH
Class II isolation equipment
Warning, consult the instruction manual
Keep in a cool, dry place
Type B equipment
Fuse
Failure to comply with the