DIGISYSTEM PS-600T Kullanım Kılavuzu
Laboratuvar Ekipmanları DIGISYSTEM PS-600T için çevrimiçi göz atın veya pdf Kullanım Kılavuzu indirin. DIGISYSTEM PS-600T 2 sayfaları. Pipette shaker
INSTRUCTION MANUAL FOR PIPETTE SHAKER PS-600T
Warnings:
* Do not remove the shaker during operation.
* Place the shaker securely away from the edge of bench to avoid the rotator mixing falling off the bench
top.
* Do not disassemble the shaker by yourself. In the event of a malfunction, the shaker should be repaired by
a qualified technician. Before repairing, turn off the power.
1. Place the pipette shaker on a flat and stable bench top. Plug it into a properly grounded outlet which
corresponds with the electrical requirements on the label back on machine.
2. Place blood diluting pipettes on the 'Holder'.
3. Select the switch model. You can place the 'Switch' at 'TIMER', 'OFF', or 'ON' position. When the 'Switch'
is at 'TIMER' or 'ON' position, the 'Power indicator' lights.
3.1 If the Timer mode is selected, you can turn the 'Timer' to a required scale. There is a bell sound to
be heard when the setting time is end.
* How to set the 'Timer' correctly?
Turn the 'Timer' over the required scale, then return the time calibration as a required scale. (For
instance, you need 5 minutes for rotation, but you need to wind to 7 or 8 minute scale at first, then
return back to 5 minute scale.)
3.2 If the 'ON' mode is selected, the rotator will rotate continuously.
3.3 If the 'OFF' mode is selected, the power will be cut off.
Maintenance and care:
Keep the machine clean all the time, so no routine maintenance is required. Cleaning can be done with a
damp cloth. Avoid the use of solvents as they may damage the machine housing.
DIGISYSTEM LABORATORY INSTRUMENTS INC.
Holder
Power
indicator
Switch
Timer
www.digisystemlab.com
REV.1-2017