Siemens syngo MR E11 Operatör El Kitabı - Sayfa 12

Endüstriyel Ekipmanlar Siemens syngo MR E11 için çevrimiçi göz atın veya pdf Operatör El Kitabı indirin. Siemens syngo MR E11 40 sayfaları. Mr system
Ayrıca Siemens syngo MR E11 için: Operatör El Kitabı (19 sayfalar)

Siemens syngo MR E11 Operatör El Kitabı
2
Safety information
12
 CAUTION
Active driver is not properly configured!
Wrong diagnosis
Ensure that the active driver is configured correctly as
described in the configuration setup (see chapter
configuration).
 CAUTION
Use of unapproved accessories or cables!
Electromagnetic interference may lead to incorrect
operation of the equipment
Only use accessories and cables that are provided by
Resoundant.
- or -
Consult Resoundant if other cables need to be used.
 WARNING
Improper storage of driver components!
Serious injury due to projectile accidents
Driver components may not function correctly
Never place the active driver component of the system in the
examination room.
Store the active driver component in an air-conditioned
equipment room.
Store the passive driver component of the system in the
examination room.
MR Elastography | Operator Manual
Print No. MR-06020G.640.08.04.02