ACD CAL-101 Kullanım Kılavuzu - Sayfa 6
Taşınabilir Jeneratör ACD CAL-101 için çevrimiçi göz atın veya pdf Kullanım Kılavuzu indirin. ACD CAL-101 15 sayfaları. Calibration gas generator
II. START-UP
To start the generator, insert the generating cell. In the bump station
configuration, press the button protruding from the bottom of the cell.
The RED LEDs will light to indicate battery life remaining. All 10 LEDs
indicate fully charged batteries. If only one RED LED is observed, the
batteries are near depletion. During the battery test time, current is supplied
to the generating cell and the air pump is not turned on in an attempt to speed
up equilibrium time. This helps to saturate the generating cell with gas.
After the battery life indication, the pump will start to pump air through the
instrument and the GREEN LEDs will begin to flash approximately once
every 4 seconds. The flashing GREEN LEDs will indicate the time remaining
in the generating cell as follows:
10 LEDs lit indicates that 2 hours remain on the cell.
1 LED lit indicates less than 12 minutes of calibration time remain.
These are the normal operating conditions for the CAL-101. Complete
equilibrium may take an additional two to five minutes, depending upon the
type and amount of gas being generated and the operating conditions of
the instrument. In the bump station configuration the unit will automatically
shut off after thirty seconds.
When the generating cell is inserted, if all 10 green LEDs come on constant,
this indicates a cell failure. Remove the cell and wait for all LED's to turn off,
then re-insert ensuring that the cell is fully inserted correctly and properly
aligned.
If the GREEN LEDs 'ramp' up and down from max to min this indicates that
the cell life is depleted and a new generating cell should be purchased.
At any time the instrument may be turned off by removing the generating
cell. The GREEN LEDs will all light indicating cell failure and the instrument
will then shut off.
Page 4