OPERATION
Please make sure that you have already read and fully understand the WARNING section on page 1.
Making Measurements with the AC Current Probe Model MN312/MN313
Connect the black lead of the current probe to "common" and the red lead to the AC current input on
your DMM or other current measuring instrument. Select the appropriate current range (400mA AC
range). Clamp the probe around the conductor to be tested with the arrow pointed toward the load. If
the reading is less than 10mA, select the lower range until you obtain the best resolution. Read the
value display on the DMM and multiply it by the probe ratio (1000/1). (If reading = 102mA, the current
flowing through the probe is 102mA x 1000 = 102000mA AC = 102A).
For best accuracy, avoid if possible, the proximity of other conductors which may create noise.
Tips for Making Precise Measurements
When using a current probe with a meter, it is important to select the range that provides the best
resolution. Failure to do this may result in measurement errors.
Make sure that probe jaw mating surfaces are free of dust and contamination. Contaminants cause air
gaps between the jaws, increasing the phase shift between primary and secondary. It is very critical for
power measurement.
MAINTENANCE:
Warning
For maintenance use only original factory replacement parts.
To avoid electrical shock, do not attempt to perform any servicing unless you are qualified to do so.
To avoid electrical shock and/or damage to the instrument, do not get water or other foreign agents
into the probe.
Cleaning
To ensure optimum performance, it is important to keep the probe jaw mating surfaces clean at all times.
Failure to do so may result in error in readings. To clean the probe jaws, use very fine sand paper (fine
600) to avoid scratching the jaw, then gently clean with a soft oiled cloth.
REPAIR AND CALIBRATION
You must contact our Service Center for a Customer Service Authorization number (CSA#). This will
ensure that when your instrument arrives, it will be tracked and processed promptly. Please write the CSA#
on the outside of the shipping container.
NOTE: All customers must obtain a CSA# before returning any instrument.
TECHNICAL AND SALES ASSISTANCE
If you are experiencing any technical problems, or require any assistance with the proper use or
application of this instrument, please contact our technical hotline:
LIMITED WARRANTY
The current probe is warranted to the owner for a period of
purchase against defects in manufacture. This limited warranty is given by AEMC
not by the distributor from whom it was purchased. This warranty is void if the unit has been tampered
with, abused or if the defect is related to service not performed by AEMC
Please print the online Warranty Coverage Information for your records.