Crowcon Clip+ Kurulum ve Kullanım Kılavuzu - Sayfa 7
Ölçüm Cihazları Crowcon Clip+ için çevrimiçi göz atın veya pdf Kurulum ve Kullanım Kılavuzu indirin. Crowcon Clip+ 15 sayfaları. Fixed life single gas monitor
2. Additional Steps:
A self test can be performed at any time during operation. Simply press the button, wait for the low
alarm level to show and press the button once more. This will activate the self test mode allowing a
user to verify the personal monitor's functionality.
Above we outlined the most common screen configurations. However if the personal monitor has been
programmed via the IR Link or has been exposed to gas additional screens may appear
Screen 4 (If applicable): If programmed with a "User ID", after the alarm set points are displayed, a
combination of numbers and or letters will scroll across the screen. This will be a max of 2 screens with
a maximum character limit on the "User ID" of 6 characters. The "User ID" can be changed/modified via
the IR Link software
Screen 5 (if applicable): If the personal monitor has been exposed to gas exceeding the low alarm set
point a value will appear with "max" next to it. This represents the peak value (highest) that the personal
monitor has seen. After this screen there will be another screen displaying a value with (hours, days, or
months) this represents the amount of time past since the peak reading.
Screen 6 (if applicable): After the peak reading and time since screens, a screen will appear with CLP
(this stands for 'clear peak'). If the user presses the button down at while this is displayed, the peak
value on the personal monitor will be reset. Note while the value will be cleared on the display the value
will be held in the personal monitor's log. See the event log section for further details. This value can
be cleared on the next screen.
Failures/FAQ's
If the self test fails, the personal monitor emits five short beeps and flashes before displaying
"Test". Repeat the self test.
If the self test fails three (3) consecutive times the personal monitor will enter Fail Safe mode.
Please contact Crowcon or your local service centre for a replacement personal monitor
During normal operation, the battery is continuously monitored. If the battery is low for more
than (3) hours the personal monitor enters Fail Safe mode.
If the battery self test fails (5) consecutive times the LCD will display E05. In case of this error
please discontinue use and contact Crowcon or your local service centre for a replacement
personal monitor.
Issue 4 – December 2016
Page 6