1. Getting Started
1.1.
Set Up / Installation instructions
C-Test is designed to be used on a flat surface and can be mounted on either a desk or on the wall.
Each C-Test should come with the following:
1.
C-Test station
2.
Removable cylinder guide
3.
One (1) USB cover
4.
One (1) USB memory stick *Underneath the cover*
5.
One (1) Multi-region Power Supply
Note: Calibration Gas cylinders are sold separately
1.2.
Power
C-Test is powered by an internal rechargeable battery. Only charge with the power supply provided.
AC Power: Ensure the Power LED is green, insert the unit and press the bump test button
Battery Power: Press the bump test or calibration button to activate the station. Pressing either button
will immediately initiate a test. The unit should be inserted before pressing the button.
Charging the Battery: While charging the Unit LED lights will cycle to indicate charging. Once fully
charged, the Power LED will be solid green. If the battery is low, the Power LED will blink green. A
complete charge takes approximately 3 hours. A full charge lasts for at least 1000 bump test cycles.
1.3.
Installation of Calibration Gas
In order to perform tests, a bottle of certified calibration gas in required. This section is located on the
right hand side of the C-Test. To install, insert the bottle and screw until tight, once properly connected
you will be able to view a pressure reading in the pressure gage. Crowcon recommends using standard
34ltr, 58ltr or 116ltr bottles. Below are the recommended calibration gas concentrations:
H2S: 25 or 15 ppm
It is important to ensure the C-Test is set to the correct levels of gas.
Warning: Calibration bottle pressure must be less than or equal to 1000 PSI. Pressures above 1000
PSI can damage the C-Test station.
2. Testing the Monitors
Place the activated Crowcon Clip or Crowcon Clip+ monitor into the bay. Make sure that the calibration
gas matches the unit type inserted. The C-Test can test different monitors by using a multi-gas blend
calibration mix.
To begin the test, press either the bump test button or the calibration button. A bump test will take
approx twenty (20) seconds and a calibration will take one (1) minute and forty five (45) seconds.
During the test, the light will display an
Issue 1 - June 2012
CO: 100 ppm O2: 18% O2 concentration
ORANGE
colour.
Page 3