Dinel ULM-53 series Manuel - Sayfa 4
Lazer Seviyesi Dinel ULM-53 series için çevrimiçi göz atın veya pdf Manuel indirin. Dinel ULM-53 series 8 sayfaları. Ultrasonic level meters
Ayrıca Dinel ULM-53 series için: Kullanım Kılavuzu (20 sayfalar), Kullanım Kılavuzu (20 sayfalar), Manuel (6 sayfalar)
Area classification
ULM–53N–_ _–_
ULM–53Xi–01–I
ULM–53Xi–02–I
ULM–53Xi–06–I
ULM–53Xi–10–I
ULM–53Xi–20–I
Installation
Level meter is installed into the upper lid of the tank (vessel), using a fixing nut or a flange.
If installed in an open channel (sumps, reservoirs, etc.), install the level meter as closest as you can to the maximum level expected.
The front of the level meter must run in parallel to the measured level.
Emitted acoustic signal must not be affected by near objects (stiffeners, ladders, mixers, unevenness, etc.), stream of filling, air
flow, etc.
Foam on the level absorbs the acoustic wave reflection which might cause malfunction of the level meter. If possible select the
location where the foaming is as low as possible. Protect the level meter against direct sunlight.
In the case of uncertainty we recommend to consult the application with the producer.
Mounting recommendation
Recommended installation
ULM–53– 01; 02; 10
ULM–53–06
ULM–53–20
Installation distance from the tank wall
(according to EN 60079-10 and EN 60079-14)
Explosive proof – suitable for explosive areas (combustible gases or vapours)
Explosive proof – suitable for explosive areas (combustible gases or vapours)
Explosive proof – suitable for explosive areas (combustible gases or vapours)
c - Maximum reach
of the level meter
d - Minimum distance
from tank wall
d > 1/12 c
(min. 200 mm)
d > 1/8 c
(min. 200 mm)
d > 1/10 c
(min. 200 mm)
Performance for non-explosive areas
II 1/2G Ex ia IIB T5 Ga/Gb with isolating repeater (IRU–420)
the whole level meter – zone 1, front head part – zone 0
II 1/2G Ex ia IIA T5 Ga/Gb with isolating repeater (IRU–420)
the whole level meter – zone 1, front head part – zone 0
II 2G Ex ia IIA T5 Gb with isolating repeater (IRU–420)
the whole level meter – zone 1
ULM–53–01; 02; 06
ULM–53–10
ULM–53–20
a - neck height
b - neck width
m - dead zone
Possible installation through the neck
Level meter installation outside the influence of filling circulation
a < 3 b
b > 100 mm
a < 1,5 b
b > 100 mm
a < 1,5 b
b > 150 mm
ULM-53-dat-4.8