Ambix Act iv Ambula t ory Pump a nd Acce ssorie s -
Ala rms a nd Safety Fea t ures
Door Ope n
Re me dy
Ta rget volume pre -a la rm
Re me dy
Ta rget volume a chieve d
Preve nt ive ma int e na nce
Te chnica l Error
Freephone 24 hour Advice Line
The pump door or lever is open.
The pump door is installed incorrectly.
The pump door mechanism is defective.
The infusion will stop.
Close the pump door and the lever.
Remove the door and re-hang.
Contact Calea Patient Services or helpline.
Only available if the target volume is active.
Be aware that there is only 30 minutes left of the infusion
from the first alarm.
The infusion will stop.
Perform for maintenance.
Contact Calea Patient Services or helpline.