5 Calibration Process
• Connect blue gas hose (Gas is set to 92 if different gas is used go into Diagnostic
mode to update details) to EC sample chamber
• Place unit in test jig located at the EC Calibration station, lock into place.
REF: EC Service and Calibration Instructions V8
Flow meter
Cal Station
Sampling tube. Connect blue hose mouthpiece
assembly to EC mouthpiece inlet port.
EC Unit
Lid and locking mechanism
Page 4
Chamber tube
Fuel cell batch#
Fuel cell connector
Foam
Pump tubing
Pump body
Pump assembly Solenoid
Make sure white solenoid
wires, are neatly tucked
away.