Direct Healthcare Group Bure Rise & Go Kullanıcı Kılavuzu - Sayfa 8

Hareketlilik Yardımı Direct Healthcare Group Bure Rise & Go için çevrimiçi göz atın veya pdf Kullanıcı Kılavuzu indirin. Direct Healthcare Group Bure Rise & Go 12 sayfaları.
Ayrıca Direct Healthcare Group Bure Rise & Go için: Teknik Kılavuz (12 sayfalar)

Direct Healthcare Group Bure Rise & Go Kullanıcı Kılavuzu

11. Maintenance

Recommended at least once per year.
Check that
General
The walker feels firm/free from play.
The walker does not rattle during manoeuvres.
The walker is level and all of its castors are in contact with
the floor.
The walker is not dirty.
Arm rests
Arm rests are intact and clean.
Arm rest widening is functional
Handle
Grips are not damaged/dirty.
Handle adjustment is functional
Battery
KCheck that the charging cable is connected to the control
box.
Check that the remote control is connected to the control
box.
Connect or fit new remote control as required.
Check that the battery, actuators and control box are not
loose.
Height adjustment
Raising and lowering are functional.
The walker feels stable at maximum settings.
There is no play between the vertical frame and bottom
frame.
Height adjustment clamps lock.
The electric actuator fitting must be free from play.
Height adjustment via remote control is functional. The
electric actuator must run smoothly at a constant speed.
Frame parts
There is no mechanical damage.
There are no scratches.
End plugs/lower frame fitted
8
BU R E RI SE & G O
Action
Clean with lukewarm soap solution or alcohol-based
cleaning agent (no petroleum products).
Clean; replace if damaged.
Replace lock knob
Clean with lukewarm soap solution (no petroleum
products).
Replace handle grips.
Fit lock knobs/plates as required.
Fit new cable; this must always be connected to the
control box (IP65 classification).
Connect; replace with new remote control as required.
Refer to the wiring diagram in the Technical manual.
Tighten; replace with new fasteners as required
Fit new guide sleeves (in frame) or end plugs (in chrome
tubes) as required.
Tighten all lock bolts.
Replace clamps.
Tighten the attachment fitting concerned; replace bolts/
lock nuts as necessary.
Check that the battery/remote control/actuator are
connected in accordance with the wiring diagram.
Charge the battery. Contact customer services.
If damage is present, contact GATE's customer
services.
Touch up as necessary.
Fit new end plugs.
DIRECTHEALTHC ARE G RO U P.CO M