12. Technical Specification
Declaration – electromagnetic immunity – for ME EQUIPMENT and ME SYSTEMS that are not LIFE-SUPPORTING
Guidance and manufacturer's declaration – electromagnetic immunity
The MAT1210001 is intended for use in the electromagnetic environment specified below.
The customer or the user of the system should ensure that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance level
Conducted RF
3 Vr
3 Vr
ms
IEC 61000- 4-6
150 kHz to 80 MHz
3 V/m
10 V/m
Radiated RF
(Professional
Healthcare Environment)
IEC 61000-4-3
10 V/m
(Home
Healthcare Environment)
80 MHz at 2.7 GHz
NOTE 1
At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2
These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection
from structures, objects and people.
a. Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios,
amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the
electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured
field strength in the location in which the Span system is used exceeds the applicable RF compliance level above, the system should
be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as
reorienting or relocating the system.
b. Over the frequency range 150kHz to 80MHz, field strengths should be less than 3V/m.
2 0
D I R E C T H E A L T H C A R E G R O U P . C O M
Electromagnetic environment – guidance
Portable and mobile RF communications equipment
ms
should be used no closer to any part of the CT515,
including cables, than the recommended separation
distance calculated from the equation applicable to the
frequency of the transmitter
Recommended separation distance
d = 1.167√P
d = 1.167√P
80 MHz to 800 MHz
d = 2.333√P
800 MHz to 2.5 GHz
Where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended separation
distance in meters (m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey,
be less than the compliance level in each frequency
range.
b
Interference may occur in the vicinity of equipment
marked with the following symbol:
12. Technical Specification
Recommended separation distances between portable and mobile RF communications equipment and the EQUIPMENT
or SYSTEM – for ME EQUIPMENT or ME SYSTEM that are not LIFE – SUPPORTING
Recommended separation distances between portable and mobile RF communications equipment and the MAT1210001
Alternating Control Unit
The MAT1210001 is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled.
The customer or the user of the system can help prevent electromagnetic interference by maintaining a minimum distance between
portable and mobile RF communications equipment (transmitters) and the system as recommended below, according to the maximum
output power of the communications equipment.
Rated maximum output
power of transmitter (W)
0.01
0.1
1
10
should
a
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be
estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1
At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2
These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection
from structures, objects and people.
S E R V I C E M A N U A L
Separation distance according to frequency of transmitter (m)
150 KHz to 80 MHz
80 MHz to 800 MHz
d = 1.167√P
d = 1.167√P
0.117
0.117
0.369
0.369
1.167
1.167
3.689
3.689
11.667
11.667