Directions for Use
1. Place the Panacea Air Pro Mattress on the bed frame with the hoses exiting at the foot
of the bed.
2. Secure mattress to the bed deck using the black straps attached to the bottom of the
mattress. Be sure to only attach the straps to the parts of the bed that articulate and
not to the stationary parts of the frame.
3. Hang the control unit on the end of the bed using the hangers, or place on a stable
horizontal surface.
4. Ensure the air lines are not "kinked" or "twisted." Connect the mattress to the control
unit by pushing the connector into the air ports located on the side of the blower until
you hear it "click" securely into place.
5. Ensure the electrical power source, if switched, is in an OFF position, then plug in the
power cord and switch to ON. The POWER ON lamp is now illuminated.
6. Set the firmness level to the highest setting and allow approximately 60 minutes for
the mattress to fully inflate. A green light next to the Alarm Silence button will indicate
when it is fully inflated.
7. After preparing the mattress surface with appropriate linens (not included), place the
resident on the mattress.
LINEN NOTE: Deep-pocketed sheets are recommended (not included). Multiple layering
of linens or under pads beneath the resident can negatively affect the mattress's pressure
management capabilities and should be avoided unless recommended by a caregiver.
8. Decrease firmness if necessary for resident comfort, support and immersion into the
mattress for optimal pressure redistribution.
a) A common method for sizing an air mattress is to check to see if suitable pressure
is selected by sliding one hand between the air mattress and the air/ foam base
under the resident's buttocks. Users should be able to feel the space between
their hand and the resident's buttocks with the acceptable range being 25 to 40
mm (1" to 1½"). This hand-check procedure is issued by AHCPR (Agency for Health
Care Policy and Research).
b) Adjust the firmness by repeatedly pressing the firmness buttons; Soft button for
more immersion (if too much space during hand check) or Firm button for less
immersion (not enough space during hand check)
9. Once the mattress is inflated, press the Mode key several times to select one of the
following:
• Auto-Firm Mode – Blows maximum airflow through both channels in the
mattress to facilitate inflation time, and is designed to revert into a normal
operating mode after inflation has been achieved. After a 15-minute period the
control unit will default to an A/P 5-minute operating cycle. Note: This mode may
be used to stiffen the mattress surface for resident procedures. However, due
to the default change to A/P mode after a 15-minute period, it is advisable that
the pump be put in Immersion mode at maximum pressure or the control unit
powered down if the procedure is to take more than 15 minutes.
• Alternating Pressure Mode – The mattress cells are alternated according to
comfort settings and Cycle time settings set on the front control panel.
12
Directions for Use