Selecting a Condenser
13
Objective
Achromatic/aplanat
magnification
condenser
1x
2x
4x
10x to 100x
Note 1:
Depending on the type of objective, the indicated numerical aperture of the objective may not be
achieved.
For example, when an objective with an N.A. of 1.4 is used, the maximum aperture of the Abbe
condenser will be only about 65% of the objective's N.A., even when the condenser aperture
diaphragm is wide open.
Refer to the condenser operating manual for more information on universal and phase contrast
condensers.
Adjusting the Field Diaphragm
14
The field diaphragm controls the amount of
illumination falling on the area of the specimen
being viewed. Turning the field diaphragm knob
changes the size of the field diaphragm. For normal
observations, the size of the diaphragm should be
slightly wider than the boundary of the field of
view. Illuminating a broader area than necessary
will result in stray light entering the field of view,
generating flare and reducing image contrast.
Appropriate field diaphragm settings are
particularly important for photomicrography. In
general, good results will be obtained by stopping
down the field diaphragm to settings slightly wider
than the area to be reproduced within the photo
frame or monitor display.
Individual Operations
13 Selecting a Condenser
Condenser ( : Optimum,
Achromat
condenser
x
x
x
x
Note 1
x
The entire field of view may not be covered if a UW eyepiece is attached.
: Suitable, x: Not suitable)
Abbe condenser
x
x
Note 1
23
1-100x condenser
Field diaphragm knob