5. Troubleshooting Tables
|2.
Electrical system
、 i Problem
Cause
Countermeasure
The power is not being supplied.
Plug the power cord into an outlet. (p.32)
The power cord is not connected to the
The lamp does not light
microscope.
Connect the power cord.
(p.32)
when the power switch is
turned on.
The lamp has not been installed.
Install the lamp.
(p.30)
The lamp is burned out.
Replace the lamp.
(p.30)
.
,
.
Use the specified lamp (refer to the
The specified lamp is not being used.
electrical specifications on p.39).
(p.30)
The lamp is near the end of its life.
Replace the lamp.
(p.30)
The lamp flickers;
The power cord is not connected securely. | Secure the connection.
(p.32)
the brightness is unstable.
The lamp is not plugged into its socket
securely.
Insert the lamp securely into its socket.
(p.30)