Nikon Eclipse E600 Talimatlar Kılavuzu - Sayfa 36
Mikroskop Nikon Eclipse E600 için çevrimiçi göz atın veya pdf Talimatlar Kılavuzu indirin. Nikon Eclipse E600 37 sayfaları. Teaching head/multi teaching head
Ayrıca Nikon Eclipse E600 için: Talimatlar Kılavuzu (38 sayfalar), Talimatlar Kılavuzu (37 sayfalar), Talimatlar Kılavuzu (21 sayfalar)
Electrical Specifications
Ml ECLIPSE E600 (Main body of the microscope)
Input Power Supply Voltage:
Lamp Rating:
Protection Class:
Operating Environmental
Conditions:
Conforming Standards
e Product with AC 120V
+ Product with AC 230V
+ AC 100/120V + 10%
50/60Hz
Current consumption:
2.4A or less
Internal fuse rating:
125V, SA
* AC 230V + 10%
50/60Hz
Current consumption:
1.2A or less
Internal fuse rating:
250V, T3.15A
12V DC, 100W halogen lamp
(OSRAM HLX 64623 or PHILIPS 7724)
Class I
* Temperature:
0 to +40°C
* Humidity:
85% RH max., non-condensing
* Altitude:
2000 m max.
* Pollution:
Degree 2
* Installation:
Category II
* Indoor use only
+ UL Listed product
* FCC 15B Class A
satisfied
This equipment has been tested and found to comply with the limits
for a Class A digital device pursuant to Part 15 of the FCC Rules.
These limits are designed to provide reasonable protections against
harmful interference when the equipment is operated in a commercial
environment. This equipment generates, uses, and can radiate radio
frequency energy and if not installed and used in accordance with
the instruction manual, may cause harmful interference to radio
communications. Operation of this equipment in a residential area
is likely to cause harmful interference in which the user will be
required to correct the interference at this own expense.
This Class A digital apparatus meets all requirements of the Canadian
Interference-Causing Equipment Regulations.
Cet appareil numérique de la Classe A respecte toutes les exigences
du Règlement sur le matériel brouilleur du Canada.
+ GS approved product
* EU Low Voltage Directive satisfied
+ EU EMC Directive satisfied
35