Nikon NR-5500 Onarım Kılavuzu - Sayfa 15
Ölçüm Cihazları Nikon NR-5500 için çevrimiçi göz atın veya pdf Onarım Kılavuzu indirin. Nikon NR-5500 25 sayfaları. Auto refractometer
TEST PROGRAM
• Model eye alignment
Place the model eye tool of the selected power on the centering tool and align it.
Focusing : Adjust the focus by pushing or pulling the J35058 model eye attaching
tool which installs the model eye.
Alignment: While observing the alignment mark, make a rough alignment of the
model eye position by turning the centering knob of the centering tool
until the asterisk * mark is positioned near the center of the display.
Then make a fine alignment while observing the "XAL: " and " YAL: "
displays to obtain "XAL: +0.00" and "YAL: +0.00" respectively. Make
sure that "TOT: 09 3 or a similar value is also displayed.
• Measuring the model eye count
When you press the measurement button, the image rotator rotates and the count is
automatically measured. The measured count is displayed in the [SET] key. Use the
[DATA] key to change the displayed count or manually set the count.
[DATA +1] : Increments the counter by 1.
[DATA -1] : Decrements the counter by 1.
• Saving the model eye count
After establishing the count, turn DIP switch No. 5 located on the right side of the
Main body to the ON position and press the [SET] key. The count is then stored in
EEPROM and two short beeps are heards.
* Press the [INCR] key and select -18.00Dpt.. (-26.00Dpt.,-24.00Dpt.,
-22.00Dpt.,+24.00Dpt. and +26.00Dpt. are out of measurement range.
So they are not necessary to calibrate.)
Start counting at - 18.00Dpt. After the storage of the count for - 18.00Dpt. is
complete, press the [INCR] key and select - 16.00Dpt.
Then measure and store the count for - 16.00Dpt.
Continue measurement sequentially up to the measurement of +22.00Dpt.. After
storing all counts for measured model eyes, turn DIP switch No. 5 to the OFF
position.
Press the [PRINT] key, and the count values stored in the EEPROM will be printed.
AX OFFSET = 67-
FOG OFFSET = +0074 —
*** DP DATA ***
DP
COUNT
D1FF.
-26D:
0000
+0050
-24D:
0050 —
+0020
-22D:
0070
+0030
-20D:
0100
TEST -11 Cylindrical axis offset
TEST - 0A Fogging target offset
Count for -24.00Dpt.
Difference between counts for
-22.00Dpt. and -20.00Dpt.
C6 - 9)
Repair Manual