TROUBLESHOOTING

NO TEMPERATURE READING ON THE THERMOCOUPLE ELECTRODE
Check the electrode for any visual defects.
Plug the electrode into the NT2000iX
1 LED changes from red to green on the front panel of the generator. If it stays red, the thermocouple may be
damaged. Replace the electrode.
Place the electrode tip into 50 °C water to see if the temperature display changes to the water temperature.
If it does not, the thermocouple may be damaged.
NO STIMULATION FELT BY PATIENT
Is the grounding pad affixed to the patient correctly? (Refer to the grounding pad "Instructions for Use"
for proper placement.)
Is the grounding pad the correct size (16 square inches of surface area)?
Is the grounding pad cable connected to the NT2000iX RF generator?
Is the electrode connected to the correct input socket as shown by the Probe LED?
Is the electrode the correct length for the needle, for example, 10 cm electrode with a 10 cm needle?
Are thermocouple electrodes and RF needles being used?
WILL NOT REACH TEMPERATURE
Is the grounding pad affixed to the patient correctly?
Is the grounding pad cable connected to the NT2000iX RF generator?
Is the electrode the correct length for the needle, for example, 10 cm electrode with a 10 cm needle?
Is the needle placed close to a blood vessel? High blood flow may cool the needle.
Is the average impedance reading of the patient normal (200 to 800 ohms)? If not, try to reduce the
impedance by repositioning the needle/electrode.
12
This document is a guide. For complete details, always refer to the NT2000iX
RF generator and set the machine to Lesion mode. Ensure the Probe
RF Generator Operator's Manual.