6. Important information
For the best results please ensure these guidelines are adhered to:
1. The Oil check monitor is very sensitive to moisture. Ensure that the surface
of the sensor is dry and that the unit is not used in conditions of high
humidity, snow, rain or fog.
2. Extremes of temperature will also affect the results, ensure that the unit is
calibrated at the ambient temperature at which the tests will be conducted.
Normal operating temperature is between 5 °C - 25 °C.
3. Ensure that the sample taken from the engine does not pick up
contamination from airborne particles or moisture. Use a vacuum sampler if
available.
4. Treat the surface of the sensor with care as excessive scratching and
abrasion of the tracks will damage the product.
5. The unit cannot be used for fire resistant fluids (water-in-oil-emulsion).
11