10.
To determine the cause of the fault close the test program, disconnect the USB cable, and swap
over the pluggable terminal blocks so that the channel 1 & 2 pair is plugged into channel 3 & 4 and 3
& 4 go into 1 & 2. Now reconnect the USB cable and run the test program and repeat step 8. If the
faulty channel moves (in this case to CH #1) then there is a connection, cable, or load cell fault. If on
the other hand the fault doesn't move then the XPV7 circuit board is faulty.
11.
Next you should compare the A/D Settings at the bottom of the dialogue box, these numbers should
match exactly with what you have on screen. If these numbers do match then there is a hardware
fault on the XPV7 circuit board.
If these numbers don't match then contact our repair centre for further instructions on resetting these
values.