CUSTOM ULTRASONICS, INC.
M
:
AINTENANCE
Clean both the interior and exterior of the cabinet on a routine basis using a damp cloth and a
standard, hospital grade, non-abrasive surface cleaner/disinfectant. The surface of the cabinet
should be dried using a soft, clean, dry, lint-free cloth.
T
R
O
ECHARGE THE
Once the baseline pump timing is established, the humidity level of the desiccant bags within the
cabinet should be monitored on a daily basis using the humidity indicator cards available from
Custom Ultrasonics, Inc.
monitoring of the desiccant bags.
When the indicator card shows that the desiccant bags need to be changed, remove the six (6) bags
of desiccant from their holding compartment. Follow instructions on the desiccant bag for the re-
activation procedure or contact Custom Ultrasonics, Inc. at 215-364-1477 for further information or
to order new bags.
To maintain the function of the storage cabinet, an extra set of desiccant bags should be held in
reserve in a sealed container. The storage cabinet can then be operated continuously.
F
R
ILTER
EPLACEMENT
The air pump black pre-filter should be replaced once a year Part # 20010.
The air pump HEPA post filter (if equipped) should be changed every 3 months Part # 13075.
M
I
AINTAIN THE
NTEGRITY OF THE
If air is heard being released through the "relief valve" when the endoscopes are connected to the
purge manifold, contact
R
-O
E
E
RDERING
NDOSCOPE
To re-order desiccant bags, humidity indicator cards, filters and repair parts, please visit our website
at www.customultrasonics.com. In the left column click on "order replacements online" then click
on "Storage Cabinet & Accessories".
Parts may also be ordered by fax at (215) 364-8760.
C
AUTION STATEMENTS
1.
Installation:
a. When installing
minimum of 2 persons to support and mount the cabinet, otherwise, bodily injury
and physical damage to the equipment and surrounding area may occur.
b. Make certain that when bolting the cabinet to the wall:
D
:
ESICCANT
Follow instructions on the humidity indicator card for proper
:
A
IR
Custom Ultrasonics, Inc.
S
TORAGE
:
Model 20000 Endoscope Storage Cabinet,
SHEET
REVISION
5 OF 7
A
P
:
UMP
Service Department.
C
S
ABINET
UPPLIES
M
20000 E
ODEL
NDOSCOPE
S
C
TORAGE
ABINET
O
'
M
PERATOR
S
:
make certain to have a
ANUAL