Symptom
The lifting column is wobbly
The lifter moves unevenly on a plane surface
The lifting arm or adjustable legs moves
uncontrolled.
The lifter does not respond to Hand control action
/ The lifting arm or legs does not move.
Battery does not charge
Troubleshooting
Possible Cause/Action
The lifting column is not correctly positioned in
the chassis or the screw in front or back is not
properly tightened.
Assemble column and chassis, and tighten screws
properly.
The castors move unevenly because of pile and
dust.
Clean the castors. Remove hair and dirt.
Short circuit between hand control and lifter.
Clean hand control plug surfaced with alcohol to
remove grease.
Defective hand control.
Replace hand control.
The control electronics has failed.
Contact local representative for service.
Emergency stop button is activated.
Turn button clockwise to reset emergency stop.
Battery is empty.
Change or charge battery.
The lifters electronics is overheated.
Wait for it to cool down.
The lifter is overloaded.
Make sure the load does not exceed SWL.
Hand control is not plugged in properly.
Plug in Hand control properly. Hand control, plug
or cord can be broken an should be replaced.
Contact between chassis and column has no
connection. Column is not correctly mounted in
chassis.
Assemble column and chassis properly.
Defective control electronics.
Contact local representative for
Faulty battery.
Try another battery.
Faulty battery charger.
Try another charger.