4. Cleaning Instructions

Due to the handling of tubes with blood samples, it is important to clean all potential
infectious parts on a daily basis. Use clinic/laboratory approved disinfectant solution.
Never use Bleach!
Note: Always use protective gloves when cleaning equipment.
Remove chute from the unit for cleaning procedure using disinfectant solution or in
1.
auto-clave equipment.
To remove the de-capper head:
2.
Remove any carousel
Press Start to let de-capper head move outwards.
Press Stop button. This will let de-capper arm to stop in out position
Push de-capper head and turn left (90 degree counter clockwise). Se figure 5.
Pull out the head gently from arm.
The de-capper head is cleaned with disinfectant solution or in auto-clave.
Warning: The sharp needle will be exposed during removal. It presents a potential hazard
from possible contact with infected blood samples. Take all necessary precaution when
handling de-capper head or chute.
De-capper head
Clean unit and cover by using disinfectant solution on a cleaning cloth.
3.
Note: Never use spray solution on Pluggo unit! It may affect the mechanical function.
[email protected] · 965 116 521 · 902 222 275 · www.akralab.es
To remove head Push and turn left 90 degree
Figure 5: Removal of de-capper head