LEVELING
For optimum performance and patient safety, all Clinton Power
Series, Bariatric, Blood Drawing Chairs should be on a clean,
level non-slip surface. Once the chair is in the desired location,
level it by adjusting the four leveling feet on the base. Turn the
feet counter clockwise to decrease the height and clockwise to
increase the height of the leg. Make certain that the chair is level
in all directions and that it does not wobble or rock.
OPERATION
CLEAN BEFORE USING
It is important to thoroughly clean the chair according to facility standards before it is put into service.
See cleaning suggestions on page 6.
The dual columns on Clinton Power Series, Bariatric, Blood Drawing Chairs are designed to
self-adjust to level the load. It is important to return the chair to its lowest position after
each use to reset the relationship between the columns. Failure to do so could result in the chair
being unresponsive to a lift command. If this happens, lower the chair to its lowest position to reset
the column relationship.
FOOT CONTROLLER
Clinton Power Series, Bariatric, Blood Drawing Chairs are
controlled with a foot-operated, wired controller.
1. Press the minus (–) side of the controller with slight foot
pressure for Down motion.
2. Press the plus (+) side of the controller with slight foot
pressure for Up motion.
The chair may be stopped and used at any position.