Tıbbi Ekipman Clinton 84100 için çevrimiçi göz atın veya pdf Kullanıcı Kılavuzu indirin. Clinton 84100 11 sayfaları. Bariatric open base power tables combine load and force capacity: 550 lbs. (249 kg)

FAQs
PROBLEM: Table is squeaking.
SOLUTION: Lubricate the table at designated points using Molykote G-0052Fg grease.
See drawing below for lubrication locations.
Back Rest
End Plug
PROBLEM: Backrest does not function properly.
SOLUTION: If the headrest does not remain flat in
the down position, loosen nut "A" and
tighten nut "B."
If headrest does not move at all, loosen
nut "B" and tighten nut "A."
After proper function has been achieved,
tighten both nuts.
10
Top Frame
Top Frame Axle
Center Axle
Lubricate using
Molykote G-0052Fg
grease.
Base Axle
Base
Back Rest Actuator Arm
Back Rest Cylinder
Motor
Slide Block, 4 total (Base and Top Frame)
Lubricate using Molykote G-0052Fg
grease, top and bottom contact points.
Nut A
Foot Lift Actuator
Arm.
DO NOT LUBRICATE.
Nut B