Charging
The charger can be placed on the wall or a table
and connected to the power outlet. The Charger
must be placed or installed in a way that makes
it easy to disconnect mains cable. The charging
light on the charger gives a yellow light when the
charger is ready for use and there is no hand con-
trol unit placed in the charger.
Battery light
Charger
Charging light
Mains cable
Description of charging light
Charging light
Yellow
Yellow
Orange
Green/yellow
Green
Orange/green
Place the hand control in the charger, make sure
the cable is connected to the lifter. After a few sec-
onds the charger light will change from yellow to
orange, indicating that the battery is charging. If
the charging light gives a constant green light, this
means the batteries are fully charged. The lifter can
not be used while charging the batteries.
If the emergency stop button is activated the
batteries will not charge.
Operating buttons
up/down
Hand control
Mode
Ready for use, no bat-
tery connected
Initialization
Fast charge
Top-off charge
Trickle charge
Error
The battery will not achieve full capacity be-
fore it has been fully charged and discharged
a couple of times.
The lifter can be operated without hand control.
The lifter is then equipped with a special charger
with a contact to be plugged in the hand control
contact on the lifter when charging. Both chargers
have the same functions.
Electrical system
On the front right side of the lifter there is a service
light which will give a signal when the lifter needs
service. When there is a green light, the lifter can
be used normally. If there is no light, push one of
the up/down buttons to activate the electronics,
and the service lamp will illuminate.
The electrical system has a power save function
which will turn off the electrical system after ten
minutes without activity. All lights will turn off. The
system is activated when pushing one of the oper-
ating buttons.
Service light
If the lifter is run up and down constantly for a
longer period, the electronics may be overheat-
ed, and then the lifter will stop. The service diode
will give a yellow flashing signal until the lifter has
cooled down sufficiently to work again.
If the lifting belt is untight, the lifter will stop to
avoid the belt from rolling in incorrectly. Run lifter
up to tighten belt.
Battery light