Maintenance

Maintaining the lift involves pre-use checklist,
cleaning, inspections and servicing. Inspect the lift
using the pre-use checklist before using the lift to
locate any faults or problems. You should also carry
out routine inspections of the lift, namely a month-
ly inspection and a periodic/yearly inspection.

Cleaning

Use
pH-neutral
disinfection is required, use isopropyl
alcohol. Do not use abrasive or corrosive
substances for cleaning.
• Cleaning shall be done on a regular basis. Clean
the outside of the lift using a slightly damp
cloth and a suitable cleaning agent.
• Remove any hair and fluff from the wheels and
make sure the wheels move freely.
• You may need to carefully clean the hand
control contact and cord with spirit to remove
any grease.

Monthly inspection

The lift must be set up for use during the
inspection.
• Inspect the lift using the pre-use checklist of
things to do before using the lift (page 3).
• Clean the lift.
• Make sure that the hand control contacts and
charger are securely in place and that the cables
on the hand control and charger are intact.
• Run the lift up and down to make sure that it
is working correctly and not making unusual
noises.
• In the event of any faults or problems, take the
lift out of operation, mark it with "out of order".
Contact service personnel.
Periodic inspection
Molift recommends a periodic inspection of the
lift once a year. Current guidelines in this area may
need to be taken into consideration. This periodic
inspection is carried out using the checklist on
page 15.
The inspection should be conducted by authorised
personnel. The owner is responsible for making
Page 14 of 16
cleaning
agents.
If
sure that the checklist is completed and signed
each time the lift is inspected.

Servicing

A service light indicates when the lift needs
servicing. The light illuminates green to indicate
that the lift can be used as normal.
The electronics record the loads lifted and how
much the lift has been used. After a certain period
of operation a signal is given to indicate that service
is required. The service light has the following
signals:
Service light
Not illuminated
Green
Yellow
Red
Red + sound signal
Service involves replacing the lift motor and
inspection/replacement of worn parts. This must be
carried out by authorised personnel as described in
the technical manual instructions.
Authorised service personnel can use the Molift
Service Tool to take readings of the lift's data
and to log servicing work. The owner of the lift is
responsible for logging servicing work and repairs.
BM08101 Molift MOVER 180 English - Rev C 19.12.2013
Battery light
Service light
Electrical
emergency
lowering
function
Status
Power-save mode
Ready for use
Book service, lift still
functional
Service required
Service urgently re-
quired