TROUBLESHOOTING:
Problem:
Stool seat cannot be raised or lowered -OR-
Stool feels sluggish or jerky when the seat is
raised or lowered.
Air is leaking from the gas cylinder.

SYMBOL DEFINITIONS:

Consult Instructions For Use
Caution, consult
accompanying documents
Manufacturer
Part Number
Authorized European
Representative
Serial Number
Never Sit On The Edge Of the
Stool Seat. Always Sit In The
Center Of The Seat.
10
ADC-09CF / ADC-09S
Correction:
Disassemble Gas Cylinder from Stool (refer to page 6 for disassembly
instructions) and inspect Cylinder for damage or excessive dirt buildup
around Cylinder Piston. Replace Cylinder if necessary.
Remove and replace Gas Cylinder. Refer to Parts List, page 9 for part
number and then contact Aseptico to reorder.