3 Installation

Mount the sensor in the mounting lug at the top of the sensor, filter facing down. The offset from vertical must
o
be max. 15
(see the illustration). DOL 53 must be mounted so that it is accessible in connection with mainte-
nance and must not be subjected to splash water.
Ensure free air passage around the sensor filter. Do not expose it to direct sunlight as this will affect the mea-
surements.
DOL 53 must warm up for half an hour before being ready to measure (see the section of LED indication).
The sensor must be placed in middle of the livestock house or in the center between walls/filters when used in
an air cleaner. Make sure there is an adequate distance to walls and air inlets.
DOL 53
140284 DOL 53 cable with connector (2 m)
2 meter wire with plug for DOL 53.
140285 DOL 53 test equipment gas, complete
Test equipment for NH3 measurement with DOL 51 / DOL 53.
140299 DOL 53 – test gas adapter