MODEL IXLdp TRANSDUCER
INSTRUCTION SHEET
WARNING! READ
zero pressure offsets that are encountered
can be adjusted using the zero adjust poten-
BEFORE INSTALLATION
tiometer. (See the Calibration section for more
details on the zeroing procedure.)
1. GENERAL:
PIPING
A failure resulting in injury or damage
The "high" and "low" pressure connection ports
may be caused by excessive overpres-
are plugged to avoid debris entering the unit.
sure, excessive vibration or pressure pul-
The plugs should be left in place until the tub-
sation, excessive instrument temperature,
ing and fittings are connected. The two
corrosion of the pressure containing parts,
pressure connections should be sealed to the
or other misuse. Consult Ashcroft Inc.,
transducer housing using teflon tape. The use
Stratford, Connecticut, USA before installing
of a dope-type sealant should not be used
if there are any questions or concerns.
since it may cause measurement errors
because of outgassing.
2. OVERPRESSURE:
Pressure spikes in excess of the rated over-
WIRING
pressure capability of the transducer may
Voltage Output:
cause irreversible electrical and/or
The IXLdp requires 12-36 VDC excitation for
mechanical damage to the pressure mea-
operation and will typically draw less than
suring and containing elements.
5mA. The warm-up is typically less than 15
seconds.
3. STATIC ELECTRICAL CHARGES:
Any electrical device may be susceptible to
Current Output without FM Approval:
damage when exposed to static electrical
The voltage required for a 4-20mA output is
charges. To avoid damage to the transducer
dependent upon the loop resistance of the cir-
observe the following:
cuit (see Fig. 2). The voltage required is pro-
• Ground the body of the transducer BEFORE
portional to the load (loop resistance) being
making any electrical connections.
driven. Figure 2 shows the minimum supply
voltage (V
) required for a given loop resis-
• When disconnecting, remove the ground
min
tance. The warm-up time is typically less than
LAST!
15 seconds.
Note: The shield and drain wire in the cable
Current Output with FM Approval:
(if supplied) is not connected to the trans-
IXKdps with the option FM approval must be
ducer body, and is not a suitable ground.
wired using the schematic diagram, 71B241,
DESCRIPTION
found on page 3 of the Installation and
The Industrial Low Pressure Differential
Maintenance instructions.
Transducer consists of a silicon diaphragm
The field wiring terminals can be accessed by
supported between two layers of metallized
unscrewing the four cover screws and remov-
glass. The Si-Glas technology combines the
ing the terminal block access cover (see Fig.
inherent high sensitivity of a variable capaci-
1). Once the cover is removed, make sure no
tance trans ducer using a micro-machined, sin-
contaminants, (e.g. water, oil, chemicals,
gle crystal diaphragm which provides excellent
grease, dirt, etc.), enter the inside of the
stability and repeatability.
enclosure. Feed the cable from the conduit
The transducer should be used with clean, dry
through one of the conduit attachment holes
air or other dry non-corrosive gases. Both
into the terminal area. Connect the conduit to
unidirectional (e.g. 0/1.0 in. W.C.) and bidirec-
the conduit connection threaded hole on the
tional (e.g. +/– 5.0 in. W.C.) pressure ranges
side of the transducer. Attach the cable wires
are offered as well as a wide selection of out-
to the appropriate terminals. The unused con-
put signals.
nection hole should be closed with a suitable
The storage temperature limits of the trans-
conduit plug.
ducer are –40 to 210°F. The unit can operate
The transducer should be wired with a multi-
between –20 and 185°F and is temperature
conductor shielded cable. Figures 3 and 4
compensated between 0 and 160°F.
show how the current and voltage output
For units with FM approval Please see
transducers should be wired. The transducer
schematic diagram 71B241.
housing should be earth grounded at the
ground screw using one of the cable's conduc-
MOUNTING
tors as shown in Figures 3 and 4.
The unit should be mounted with #8 or #10
screws using the three mounting feet provided
When the cover is reinstalled, make sure the
(see Fig. 1). Easy access to the covers may
gasket is seated correctly and all four screws
be a consideration when mounting. The trans-
are properly engaged and tightened.
ducer can be mounted in any orientation with
virtually no effect on calibration. Any minor
© 2015 Ashcroft Inc., 250 East Main St., Stratford, CT 06614-5145, U.S.A. Tel: 203-378-8281, Fax: 203-385-0499, email: [email protected], www.ashcroft.com
All sales subject to standard terms and conditions of sale. I&M011-10062 (250-2985) Rev. I 06/15
CALIBRATION
The zero adjustment for the transducer can be
accessed by unscrewing the four cover screws
and removing the zero pot access cover (see
Fig. 1). Once the cover is removed, make sure
no contaminants, (e.g. water, oil, chemicals,
grease, dirt, etc.), enter the enclosure. The
zero adjustment range is approximately ±10%
of span.
1
⁄
˝ NPT
4
The offset or zero adjustment potentiometer is
shown in Figure 1. A hole is provided in the
fiber insulation board to access the pot adjust-
ment screw. Before you adjust, make sure a
short tube is connected from the "low" port to
the "high" port of the transducer. This connec-
tion will ensure that both ports are at the same
pressure. A clockwise rotation raises the output.
DO NOT ADJUST THE SPAN
ADJUSTMENT POTENTIOMETER
Proper span calibration requires a pressure
standard three to five times more accurate
than the accuracy of the transducer.
When the cover is reinstalled, make sure the
gasket is seated properly and all four screws
are properly engaged and tightened.
SERVICE
There are no user serviceable parts inside the
transducer. Servicing should only be carried
out by factory-trained personnel. Service prob-
lems encountered in the field might be
mechanical or electrical in nature.
Mechanical problems might be attributed to
plugged or leaking pressure lines, or faulty
pressure sources.
Electrical problems might be attributed to
improper wiring, bad connections (e.g. open,
shorts, etc.), malfunctioning or improper power
supply.
PHYSICAL
Weight: approx. 2 lbs.
NEMA 4X rated enclosure
OPTIONAL FEATURES
Variable Damping
A third potentiometer, placed to the left of
the zero potentiometer, allows the user to
adjust the unit's response time.
Multiple Range
By relocating the shunt jumper, the user can
"re-range" the unit. The unit can be re-
ranged by a
1
⁄
,
1
⁄
,
1
⁄
and a
1
⁄
. For example, a
2
3
4
5
unidirectional 5 in. W.C. unit can become a
2.5 in. W.C., 1.67 in. W.C., 1.25 in. W.C. or a
1.0 in. W.C. range.