Temperature
Figure 3.1: Typical transducer placement in a cross-flow system
Installing the transducers
Connect pressure transducers to the feed, retentate, and one of the permeate ports of your cross-
flow cartridge. Connect the transducer signal cables to the appropriate jacks—Inlet (or feed), Outlet
(or retentate and Permeate—on the back of the display unit. Plug the temperature sensor signal
cable into the jack on the back of the display unit. Place the temperature probe in the feed reservoir,
permeate collection vessel, or integrate it into the retentate line using a tee adaptor andLuer-Lok
connectors.
MidJet pressure display unit Instructions for use 18117465 AD
Tubing Pinch
Retentate
Clamp
Probe
Feed
Reservoir
Feed
Pump
CAUTION
The pressure display unit is calibrated at the factory with the
installed transducers. If you swap the transducers by plugging
them into different terminals on the back of the pressure display
unit, the accuracy of the pressure reading will be compromised. To
maintain the accuracy of the pressure display unit, do not swap the
transducers by plugging them into different terminals on the back
of the pressure display unit. Recalibrate the pressure display unit
when you install new pressure transducers.
Cross-Flow
Cartridge
Permeate Port
Blanked Off
Transducer
Pressure Display Unit
3 Setup and operation
Permeate
Line
Permeate
Collection
Vessel
Signal
Cable